CONTEC CMS6000 Patient Monitor ETCO2 Introduction:
The monitor has abundant monitoring functions and is used for monitoring adults, pediatrics,s, and neonate.
Users may select different parameter configurations according to different requirements.
It displays real-time data and waveform by 8” TFT color LCD, and up to 8-channel waveform and all monitoring parameters can be displayed simultaneously. 48mm thermal printer is optional.
In addition, it can be connected to the central monitoring system via a wired/wireless network to form a network monitoring system.
This device can monitor such parameters as ECG, RESP, SpO2, PR, NIBP and dual-channel TEMP, etc. It integrates parameter measurement modules, display, and recorder in one device to form compact and portable equipment. At the same time, its built-in replaceable battery provides convenience for patient moving.
This device can monitor such parameters as ECG, RESP, SpO2, PR, NIBP and dual-channel TEMP, etc. It integrates parameter measurement modules, display, and recorder in one device to form compact and portable equipment. At the same time, its built-in replaceable battery provides convenience for patient moving.
CONTEC CMS6000 Patient Monitor ETCO2 FUNCTIONS:
Standard parameters: ECG, RESP, SpO2, PR, NIBP, dual-channel TEMP
- ECG Heart rate(HR)
ECG waveform
Arrhythmia and ST-segment analysis - RESP Respiration rate(RR)
Respiration waveform - SpO2 Oxygen saturation(SpO2)
Plethysmogram(PLETH) waveform
Pulse rate(PR)
Bar graph - NIBP Systolic pressure(SYS), Diastolic pressure(DIA), Mean Pressure(MEAN)
- TEMP T1, T2, TD
It has abundant functions, such as audible and visual alarm, trend data storage and output, NIBP measurement, alarm event marking and drug concentration calculation, etc
CONTEC CMS6000 Patient Monitor ETCO2 FEATURES:
- 8” TFT color LCD, multi-language interface(Simplified Chinese, Traditional Chinese, English, French, German, Turkish, Spanish, Portuguese, Italian, Dutch, Romanian, Russian, Kazakhstan, Polish, Czech).
- Fanless design, quiet, energy-saving, and clean, which reduces the possibility of cross-infection.
- All-round monitor for adult, pediatric, and neonate.
- With the standard interface, OxyCRG, trend chard, big character interface, and view bed, convenient to observe.
- Waveform, parameter color, and location can be set optionally.
- Operation mode: keys and buttons.
- Display up to 8-channel waveform on one screen.
- Display 7-lead ECG waveform on one screen, and with ECG step function.
- Adopt digital SpO2 technology, anti-motion, and anti-ambient light interference, and measurement can be performed under the circumstance of weak filling.
- HRV analysis function.
- NIBP measurement mode: Manual/AUTO/STAT, storage for 4800-group NIBP data.
- Review for 71 alarm events of all parameters and 60 arrhythmia alarm events.
- Drug concentration calculation and titration table functions.
- Print trend table data with one key.
- Connect to Central Monitoring System by 3G, WiFi, or wired mode.
- AC/DC, built-in rechargeable lithium battery achieves uninterrupted monitoring.
- Anti-high frequency surgical unit, defibrillation-proof(special leads are necessary).
CONTEC CMS6000 Patient Monitor ETCO2 Performance:
- ECG
Lead mode: 3-lead or 5-lead
Lead selection: I, II, III, aVR, aVL, aVF, V
Waveform: 5-lead, 2-channel
3-lead, 1-channel
Gain: 2.5mm/mV, 5.0mm/mV, 10mm/mV, 20mm/mV, 40mm/mV
Scan speed: 12.5mm/s, 25 mm/s, 50 mm/s - HR:
Measurement and alarm range: 15~350bpm
Accuracy: ±1% or ±1bpm, whichever is greater
Alarm accuracy: ± 2 bpm
Resolution: 1 bpm
ST-segment monitoring:
Measurement and alarm range: -2.0mV~+2.0mV
Accuracy: -0.8mV~+0.8mV ±0.04mv or ±10%, whichever is greater
Other range: unspecified
Arrhythmia analysis: ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT>2, PVC, TACHY, BRADY, MISSED BEATS, PNP, PNC
Pacemaker: yes - RESPIRATION
Method: R-F(RA-LL) Impedance
Respiration rate:
Measurement and alarm range: 0~150rpm
Resolution: 1 rpm
Measurement accuracy: ±2 rpm
Alarm accuracy: ±3 rpm
Apnea alarm: 10~40s
Scan speed: 6.25 mm/s, 12.5 mm/s, 25 mm/s - NIBP
Method: Oscillometry
Mode: Manual/AUTO/STAT
Measurement interval in AUTO mode: 1/2/3/4/5/10/15/30/60/90/120/240/480/960 minutes
Measurement period in STAT mode: 5 minutes
Measurement and alarm range: 10 ~ 270 mmHg
Resolution: 1 mmHg
Cuff pressure accuracy: ±3 mmHg
Measurement accuracy:
Maximal mean deviation: ±5 mmHg
Maximal standard deviation: 8mmHg
Over-pressure protection:
Adult mode: 297±3 mmHg
Pediatric mode: 240±3 mmHg
Neonatal mode: 147±3 mmHg - SpO2
Measurement and alarm range: 0 ~ 100%
Resolution: 1%
Measurement accuracy: 70%~100%: ±2%;
0%~69%: unspecified - PR
Measurement and alarm range: 30 ~ 250 bpm
Measurement accuracy: ±2 bpm or ±2%, whichever is greater - TEMP
Channel: dual-channel
Measurement and alarm range: 0 ~ 50℃
Resolution: 0.1 ℃
Accuracy: ±0.1 ℃ - EtCO2
Method: Sidestream or Mainstream
Measurement range for CO2: 0 ~ 150mmHg
Resolution for CO2:
0.1 mm Hg: 0 ~ 69 mm Hg
0.25 mm Hg: 70 ~ 150 mm Hg
Accuracy for CO2:
0 ~ 40 mm Hg: ±2 mm Hg
41 ~ 70 mm Hg: ±5%
71 ~ 100 mm Hg: ±8%
101 ~ 150 mm Hg: ±10%
AwRR range: 2~150 rpm
AwRR accuracy: ±1 rpm
Apnea alarm: yes
IBP
Channel: dual-channel
Label: ART, PA, CVP, RAP, LAP, ICP, P1, P2
Measurement and alarm range: -10~300 mm Hg
Resolution: 1 mmHg
Accuracy: ±2% or 1mm Hg, whichever is greater
Power supply:100-240V~, 50/60Hz
Safety classification: Class I, type CF defibrillation-proof applied part
- Standard Configuration:
ECG, NIBP, SpO2, Resp, 2-Temp - Optional:
2-IBP, ETCO2,Network, Thermal Recorder
CONTEC CMS6000 Patient Monitor ETCO2 Accessories:
- Adult fingertip SpO2 probe (5-pin)
- Adult NIBP cuff
- NIBP extension tube
- ECG lead cable
- ECG electrode
- Temperature probe
- Power adapter
- Power cord
- User Manual
CONTEC CMS6000 Patient Monitor ETCO2 Physical Characteristic:
- Dimension: 235 mm(L) × 141 mm(W) × 222 mm(H)
- Weight: 2.5 Kg
To activate demo mode: Code 2088












Reviews